CheckMyRecalls

Vision RT Ltd recalls The AlignRT InBore system contains six Class 1…

Vision RT Ltd recalls The AlignRT InBore system contains six Class 1 lasers which enables…

Published
14 Jan 2026
Units affected
56
Agency
FDA
Country
United States

The hazard

Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.

Brand / company
Vision RT Ltd
Codes / lots
Part Number V000629
Distribution
U.S. and OUS
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 14 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.