The hazard
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
- Brand / company
- Medica Corporation
- Codes / lots
- Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
- Distribution
- International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 25 Feb 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.