CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Instrumentation Laboratory recalls GEM Premier 5000 PAK and 1 more

Read the official FDA notice

Published
18 Mar 2026
Units affected
20 products
Agency
FDA
Country
United States

The hazard

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GEM Premier 5000 PAK, Part No. 00055360004.

GEM Premier 5000; Part No. 00055415005.

GEM Premier 5000; Model No. 00055430010.

GEM Premier 5000; Part No. 00055430008.

GEM Premier 5000; Part No. 00055415008.

GEM Premier 5000 PAK; Part No. 00055407511.

GEM Premier 5000 PAK; Part No. 00055360011.

GEM Premier 5000 PAK; Part No. 00055407508.

GEM Premier 5000; Part No. 00055445008.

GEM Premier 5000 PAK; Part No. 00055407510.

GEM Premier 5000 PAK; Part No. 00055360008.

GEM Premier 5000; Part No. 00055445004.

Brand / company
Instrumentation Laboratory
Codes / lots
Part number: 00055360004; UDI: 08426950899662; Lot No. 251113U, 251201AI, 251202Q, 251223E, 251223P, 251229X. | Part number: 00055415005; UDI: 08426950807667; Lots No. 251125S, 251201AJ, 251212O, 251218V, 251229X. | Part number: 00055430010; UDI: 08426950807759; Lots No.: 251021K, 251027AA, 251110AH, 251114K, 251114O, 251114P, 251114Y, 251117AA, 251117AC, 251117AG, 251117AM, 251117AN, 251117P, 251117T, 251117V, 251117Y, 251118A, 251118B, 251118R, 251118X, 251119A, 251120R, 251120U, 251121Q, 251124O, 251124Q, 251125I, 251126S, 251126T, 251126U, 251201W, 251201X, 251202G, 251202H, 251203L, 251204R, 251205L, 251208N, 251211W, 251215D, 251215Z, 251216O, 251216S, 251217L, 251217V, 251218AB, 251218B, 251218X, 251218Y, 251219A, 251219O, 251222AE, 251222E, 251222F, 251222K, 251222M, 251222P, 251222Z, 251223A, 251223B, 251223C, 251223I, 251223M, 251223O, 251229D, 251229E, 251229I, 251229S, 251231B, 251231G. | Part number: 00055430008; UDI: 08426950807735; Lots No.: 251106X, 251110AF, 251202P, 251203Q, 251215J, 251229AG, 251229X, 251230P. | Part number: 00055415008; UDI: 08426950807674; Lots No. 251021P, 251106X, 251110AF, 251111O, 251113J, 251113V, 251117AD, 251118D, 251124Z, 251126Q, 25120
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 18 Mar 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.