The hazard
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
- Brand / company
- Siemens Healthcare Diagnostics, Inc.
- Codes / lots
- Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Mar 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.