Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment

Read the official FDA notice

Published
25 Mar 2026
Units affected
1098 units
Agency
FDA
Country
United States

The hazard

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Brand / company
Howmedica Osteonics Corp.
Codes / lots
Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 25 Mar 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.