The hazard
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Radius VSM ECG pre-connected Set, REF: 4842, 4695
Radius VSM Disposable NIBP Cuff, REF: 4825, 4826
- Brand / company
- Masimo Corporation
- Codes / lots
- REF/UDI-DI/Lot: 4842/10843997011003/25AJH, 25KMA, 25MSC 4695/10843997016015/24N17, 25BBJ, 25DMH, 25FBR, 24KLV, 24N17, 25AGZ, 25BBJ, 25D10, 25DMH, 25FBR, 25GJM, 25KBF | REF/UDI-DI/Lot: 4825/00843997006989/21G6S, 22JHV, 23H03, 23J05, 24DBV, 24NGG, 25ADG, E21D31, E21JWM, E23FVD; 4826/10843997006993/21HVX, 22ELB, 23G31, 23H04, 23HWL, 23J06, 24DBW, 24MNP, 24NGF, 25ADF, E21D32, E21F04, E21JFA, E21KHK
- Distribution
- Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 8 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.