CheckMyRecalls

Hologic, Inc recalls Brand Name

Published
6 May 2026
Units affected
1200 units
Agency
FDA
Country
United States

The hazard

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Brand / company
Hologic, Inc
Codes / lots
Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
Distribution
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 6 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.