CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model…

Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000

Read the official FDA notice

Published
8 Apr 2026
Units affected
12
Agency
FDA
Country
United States

The hazard

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Brand / company
Reflexion Medical, Inc.
Codes / lots
Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 8 Apr 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.