Beta Bionics, Inc. recalls iLet Ace Pump Kit REF
- Published
- 22 Apr 2026
- Units affected
- 15 units
- Agency
- FDA
- Country
- United States
The hazard
It was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Brand / company
- Beta Bionics, Inc.
- Codes / lots
- ILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Apr 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.