Medtronic, Inc. recalls Medtronic Sphere-9 Catheter

Read the official FDA notice

Published
6 May 2026
Units affected
120698 units
Agency
FDA
Country
United States

The hazard

Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.

Brand / company
Medtronic, Inc.
Codes / lots
US: Basic UDI Number 0763000B000262284, GTIN: 00763000871741; 00763000973391; 00763000973506; 00763000993269; 00199150003709. OUS: 00763000871734; 00763000973384; 00763000973513; 00763000811341; 00763000911515; 00763000963675; 00199150003716. Serial Numbers: 227076469, 2301234000, 2301321000, 2301373000, 2301543000, 2301547000, 2301551000, 2301552000, 2301556000, 2301558000, 2301559000, 2301560000, 2301709000, 0012573781, 0012576276, 0012578858, 0012583026, 0012583027, 0012590913, 0012590914, 0012594540, 0012594541, 0012609253, 0012609255, 0012609258, 0012614806, 0012614807, 0012614808, 0012619893, 0012619896, 0012619898, 0012619899, 0012623987, 0012623988, 0012623991, 0012631219, 0012631220, 0012631221, 0012631222, 0012635279, 0012635281, 0012635282, 0012635283, 0012635284, 0012641293, 0012641294, 0012641295, 0012641297, 0012643392, 0012643393, 0012643396, 0012643399, 0012645205, 0012645206, 0012645208, 0012655005, 0012655006, 0012655008, 0012655009, 0012655010, 0012655011, 0012655012, 0012655013, 0012655014, 0012655015, 0012655016, 0012655017, 0012665669, 0012665670, 0012665671, 0012665673, 0012665674, 0012665675, 0012670372, 0012670375, 0012670377, 0012670380, 0012670381, 001267
Distribution
Worldwide
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 6 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.