CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Boston Scientific Corporation recalls LUX-DX II and 1 more

Read the official FDA notice

Published
13 May 2026
Units affected
2 products
Agency
FDA
Country
United States

The hazard

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5

Brand / company
Boston Scientific Corporation
Codes / lots
LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787 | LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284 145336 145684 146709 147185 147618 148972 149028 153330 153628 1551
Distribution
Worldwide distribution - US Nationwide and the country of Belgium.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 13 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.