CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25…

Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System

Read the official FDA notice

Published
6 May 2026
Units affected
71725 (38025 US, 33700 OUS)
Agency
FDA
Country
United States

The hazard

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Brand / company
Siemens Healthcare Diagnostics Inc
Codes / lots
Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
Distribution
Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 6 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.