The hazard
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
- Brand / company
- Centinel Spine, Inc.
- Codes / lots
- Model Number: PDSXL5; UDI-DI: 00843193113979; Lot Number: 2026-0026 | Model Number: PDSXL6; UDI-DI: 00843193113986; Lot Number: 2026-0027
- Distribution
- US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 13 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.