Abiomed, Inc. recalls Automated Impella Controller (AIC)
- Published
- 3 Jun 2026
- Units affected
- 91 units
- Agency
- FDA
- Country
- United States
The hazard
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Automated Impella Controller (AIC); Product Code: 0042-0000-US;
- Brand / company
- Abiomed, Inc.
- Codes / lots
- Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 3 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.