CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Sagemax Bioceramics, Inc. recalls NexxZr T Multi A3.5 W98-16mm

Read the official FDA notice

Published
17 Jun 2026
Units affected
113
Agency
FDA
Country
United States

The hazard

Yttrium-stabilized zirconium oxide discs for fixed all-ceramic dental restorations may have reduced or missing dentin layer and when used clinically in the patient's mouth there is an increased risk of fractures under functional stress.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

NexxZr T Multi A3.5 W98-16mm, REF: 745451

Brand / company
Sagemax Bioceramics, Inc.
Codes / lots
UDI-DI: 00842271175373, Lot: YBDGLD
Distribution
Worldwide - US Nationwide distribution in the states of CA, MO, MN, NY, FL, ID and the countries of AT, CZ, DE, EE, EG, FR, GR, ID, IT, KZ, RS, SE, SI, TZ, VN.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 17 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.