CheckMyRecalls

The Binding Site Group, Ltd. recalls Brand Name

Published
17 Jun 2026
Units affected
1
Agency
FDA
Country
United States

The hazard

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Brand / company
The Binding Site Group, Ltd.
Codes / lots
Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.
Distribution
US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 17 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.