The hazard
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
- Brand / company
- Fresenius Kabi USA, LLC
- Codes / lots
- Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
- Distribution
- Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 10 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.