PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M15…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M15 and 7 more

Read the official FDA notice

Published
10 Jun 2026
Units affected
8 products
Agency
FDA
Country
United States

The hazard

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
UDI (01)00884838085282(21)209, (01)00884838099210(21)100, (01)00884838099210(21)254, (01)00884838099210(21)256, (01)00884838099210(21)257, (01)00884838099210(21)265, (01)00884838099210(21)255, (01)00884838099210(21)99, (01)00884838085282(21)260, (01)00884838085282(21)261, (01)00884838099210(21)69, (01)00884838099210(21)12, (01)00884838085282(21)314, (01)00884838085282(21)259, (01)00884838099210(21)15, (01)00884838099210(21)19, (01)00884838099210(21)23, (01)00884838099210(21)17, (01)00884838085282(21)258, (01)00884838099210(21)55, (01)00884838099210(21)45, (01)00884838085282(21)270, (01)00884838099210(21)187, (01)00884838099210(21)223, (01)00884838099210(21)44, (01)00884838099210(21)113, (01)00884838099210(21)112. | UDI (01)00884838099203(21)3, (01)00884838099203(21)7, (01)00884838085275(21)131, (01)00884838099203(21)54, (01)00884838085275(21)84, (01)00884838099203(21)6, (01)00884838099203(21)93, (01)00884838085275(21)132, (01)00884838099203(21)69, (01)00884838099203(21)9, (01)00884838099203(21)115, (01)00884838099203(21)85, (01)00884838099203(21)19, (01)00884838099203(21)49, (01)00884838099203(21)108, (01)00884838085275(21)108. | UDI (01)00884838099265(21)104, (01)00884838085350(21
Distribution
US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 10 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.