Hamilton Medical AG recalls Breathing circuit set

Read the official FDA notice

Published
8 Jul 2026
Units affected
27,270 pcs
Agency
FDA
Country
United States

The hazard

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Brand / company
Hamilton Medical AG
Codes / lots
REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.
Recall status
Ongoing