CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

GE Medical Systems SCS recalls GE Healthcare AW Server 3.2 ext.6.5

Read the official FDA notice

Published
1 Jul 2026
Units affected
340 servers
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

Brand / company
GE Medical Systems SCS
Codes / lots
REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.