Innovasis, Inc recalls Anti-Torque
- Published
- 12 Aug 2026
- Units affected
- 5
- Agency
- FDA
- Country
- United States
The hazard
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
- Brand / company
- Innovasis, Inc
- Codes / lots
- LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014
- Distribution
- US Nationwide distribution in the states of TX, CA.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 12 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.