The hazard
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
ICU Medical Plum Solo Precision IV Pump, v1.0.3, Item Number 400010401; infusion pump
- Brand / company
- ICU Medical, Inc.
- Codes / lots
- 1) UDI/DI +M335400021, Serial Numbers: 50003704, 50003710, 50004046, 50004072, 50004149, 50004154, 50004156, 50004163, 50004166, 50004168, 50004177, 50004183, 50004205, 50004218, 50004220, 50004249, 50004258, 50004262, 50004263, 50004264, 50004271, 50004300, 50004303, 50004304, 50004305, 50004307, 50004308, 50004310, 50004312, 50004314, 50004316, 50004318, 50004319, 50004323, 50004324, 50004325, 50004326, 50004328, 50004330, 50004331, 50004332, 50004333, 50004334, 50004335, 50004336, 50004338, 50004339, 50004340, 50004341, 50004342, 50004344, 50004345, 50004346, 50004347, 50004348, 50004349, 50004350, 50004351, 50004352, 50004356, 50004358, 50004359, 50004361, 50004363, 50004365, 50004366, 50004370, 50004372, 50004373, 50004374, 50004376, 50004380, 50004382, 50004383, 50004384, 50004385, 50004386, 50004387, 50004389, 50004390, 50004391, 50004392, 50004393, 50004394, 50004395, 50004396, 50004398, 50004399, 50004401, 50004403, 50004405, 50004407, 50004409, 50004410, 50004411, 50004413, 50004415, 50004416, 50004417, 50004424, 50004425, 50004426, 50004428, 50004431, 50004433, 50004435, 50004436, 50004438, 50004439, 50004442, 50004443, 50004444, 50004445, 50004448, 50004449, 50004451, 5
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.