CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

CMR SURGICAL LIMITED recalls Brand Name

Read the official FDA notice

Published
15 Jul 2026
Units affected
2
Agency
FDA
Country
United States

The hazard

Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1 Product Description: The Versius Surgical System is a modular, open console, software controlled, electromechanical system designed for surgeons to perform minimally invasive surgery.

Brand / company
CMR SURGICAL LIMITED
Codes / lots
UDI-DI 05060548211226 Model No V-SS-1020 Serial Numbers 66495, 66496 Software Revision Version 62.1 There is no expiry date for this device.
Distribution
US Nationwide distribution in the state of California.