CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

GE Medical Systems Information Technologies Inc recalls GE Healthcare…

GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360

Read the official FDA notice

Published
12 Aug 2026
Units affected
163 units
Agency
FDA
Country
United States

The hazard

Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

Brand / company
GE Medical Systems Information Technologies Inc
Codes / lots
GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU219400034SA, VU219400045SA, VU219400055SA, VU220270027SA, VU220510048SA, VU220510078SA, VU221100161SA, VU221100181SA, VU221150116SA, VU222120212SA, VU222120220SA, VU222120239SA, VU222120243SA, VU222120249SA, VU222120250SA, VU222120253SA, VU222120258SA, VU222120260SA, VU222120261SA, VU222120266SA, VU222120269SA, VU222140004SA, VU222200296SA, VU222520114SA, VU223120014SA, VU223120041SA, VU223420119SA, VU224050158SA, VU224070091SA, VU224200028SA, VU224420159SA, VU224520140SA, VU225170260SA, VU225420050SA, VU225520037SA, VU226040025SA, VU419320042SA, VU421490135SA, VU421490151SA, VU421490153SA, VU421490155SA, VU421490159SA, VU422460150SA, VU422460194SA, VU422460203SA, VU422460303SA, VU422480164SA, VU422500239SA, VU422500240SA, VU422500241SA, VU422500242SA, VU422500244SA, VU422500245SA, VU422500246SA, VU422500247SA, VU422500248SA, VU422500249SA, VU422500250SA, VU422500251SA, VU422500254SA, VU422500260SA, VU422500261SA, VU422500264SA, VU422500265SA, VU422500266SA, VU422500267SA, VU422500269SA, VU422510038SA, VU422510049SA, VU422510060SA, VU422510061SA, VU423050080SA, VU423050
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 12 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.