The hazard
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
- Brand / company
- KENT IMAGING, INC.
- Codes / lots
- UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
- Distribution
- US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
- Recall status
- Ongoing