CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

KENT IMAGING, INC. recalls SnapshotGLO Wound Imaging Device

Read the official FDA notice

Published
29 Jul 2026
Units affected
77
Agency
FDA
Country
United States

The hazard

The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.

Brand / company
KENT IMAGING, INC.
Codes / lots
UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 29 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.