CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Lutronic Corporation recalls Product Name

Read the official FDA notice

Published
16 Sept 2026
Units affected
1065
Agency
FDA
Country
United States

The hazard

Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Brand / company
Lutronic Corporation
Codes / lots
Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV026B015, DV026B016, DV026B017, DV026B018, DV026C006, DV026C007, DV026C008, DV026C009, DV026C010, DV025M035
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and European Union.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 16 Sept 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.