The hazard
Momentary absence of On-Screen "X-Ray On" banner during run readiness check.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Lago X
Ami HTX
- Brand / company
- Spectral Instruments Inc
- Codes / lots
- All serial numbers. | All serial numbers.
- Distribution
- United States.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 26 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.