Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw
- Published
- 7 Oct 2026
- Units affected
- 34
- Agency
- FDA
- Country
- United States
The hazard
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
- Brand / company
- Wright Medical Technology, Inc.
- Codes / lots
- UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026
- Distribution
- US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 7 Oct 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.