TECHNO-PATH MANUFACTURING LTD. recalls Techno-Path Multichem AE…
TECHNO-PATH MANUFACTURING LTD. recalls Techno-Path Multichem AE Unassayed Level 2 REF and 2 more
- Published
- 28 Oct 2026
- Units affected
- 3 products
- Agency
- FDA
- Country
- United States
The hazard
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
- Brand / company
- TECHNO-PATH MANUFACTURING LTD.
- Codes / lots
- UDI:(01)05391523442817/Lot # 65705262 | UDI: (01)05391523442800/Lot # 65705261 | UDI:(01)05391523442824/-Lot # 65705263
- Distribution
- US Nationwide distribution in the state of Massachusetts.
- Recall status
- Ongoing