The hazard
Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122008NA Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122009 Product Description: LIKORALL 242 S, NATURAL Product Code: 3122009NA Product Description: LIKORALL 242 S, NATURAL Product Code: 3122010 Product Description: LIKORALL 242 S, WHITE Product Code: 3122011 Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122011NA Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122012 Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122012NA Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122501 Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122501NA Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122502 Product Description: LIKORALL 250 ES, WHITE Product Code: 3122502NA Product Description: LIKORALL 250 ES, WHITE Product Code: 3123001 Product Description: LIKORALL 243 ES Product Code: 3123001NA Product Description: LIKORALL 243 ES Product Code: 3123002 Product Description: LIKORALL 243 ES, WHITE Product Code: 3123002NA Product Description: LIKORALL 243 ES, WHITE Product Code: 3124050 Product Description: LIKORALL 250 S, IRC
- Brand / company
- Baxter Healthcare Corporation
- Codes / lots
- Product Code: 3121001 Product Description: LIKORALL 200 Lot/Serial Number: Intervals: 8400054-8450499 and 9000000-9000999 UDI-DI Number: 00887761981911 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Greater than or equal to 8104548 UDI-DI Number: 00887761979710 Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Interval: 429087-431032 UDI-DI Number: 00887761979727 Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Greater than or equal to: 8101715 UDI-DI Number: 00887761979734 Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Interval: 429186-431035 UDI-DI Number: 00887761979741 Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Lot/Serial Number: Greater than or equal to: 8104481 UDI-DI Number: 00887761979758 Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Lot/Serial Number: Interval: 429041-430984 UDI-DI Number: 00887761979765 Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Lot/Serial Number: Greater than or equal to: 8100815 UDI-DI Number: 00887761979772 Product Co
- Distribution
- Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, and Wyoming. International distribution in pending.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 21 Oct 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.