The hazard
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
- Brand / company
- Medline Industries, LP
- Codes / lots
- MDR107000P UDI-DI 00080196288284 All lots | MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots
- Distribution
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 7 Oct 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.