The hazard
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
- Brand / company
- Beckman Coulter, Inc.
- Codes / lots
- Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial Numbers | Product/UDI-DI: IQ200 SPRINT/10837461001256, IQ200 SELECT/10837461001300, IQ200 ELITE/10837461001355. All Serial Numbers
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
- Recall status
- Ongoing