CheckMyRecalls

Stryker Leibinger GmbH & Co. KG recalls Precision Pointer…

Stryker Leibinger GmbH & Co. KG recalls Precision Pointer Electromagnetic (8000-050-001)

Published
4 Nov 2026
Units affected
513
Agency
FDA
Country
United States

The hazard

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Precision Pointer Electromagnetic (8000-050-001)

Brand / company
Stryker Leibinger GmbH & Co. KG
Codes / lots
Lot 7000413361 UDI-DI 07613327413205
Distribution
Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 4 Nov 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.