Abbott Medical recalls
4 recall notices naming Abbott Medical since January 2025, in United States. The most recent was published 17 Jun 2026.
Latest recalls (4)
- Abbott Medical recalls Merlin" Patient Care System (PCS) software Merlin" PCS 3650 program…FDASoftware malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command…
- Abbott Medical recalls Eterna SCS IPG (Implantable Pulse Generator) and 1 moreFDAThere is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain…
- Abbott Medical recalls Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI…FDADue to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery…
- Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer…FDADue to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a…