CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

Abbott Medical recalls Merlin" Patient Care System (PCS) software…

Abbott Medical recalls Merlin" Patient Care System (PCS) software Merlin" PCS 3650 program…

Read the official FDA notice

Published
17 Jun 2026
Units affected
11182 programs
Agency
FDA
Country
United States

The hazard

Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Brand / company
Abbott Medical
Codes / lots
Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 17 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.