CheckMyRecalls

Medical devices recalls

2,678 recalls since January 2025 from Health Canada, FDA, newest first. Last notice 19 Nov 2026.

Page 20 of 54

  1. Abiomed, Inc. recalls Automated Impella Controller (AIC)FDAPotential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation…
  2. Abiomed, Inc. recalls Purge CassetteFDAIncreased risk of purge leaks with Generation 1 purge cassettes.
  3. Intuitive Surgical, Inc. recalls Brand NameFDADue to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
  4. Burlington Medical, LLC recalls Burlington MedicalFDAPotential for attenuation degradation over time, decreasing the lifespan.
  5. Ultradent Products, Inc. recalls Jiffy Original Composite Polisher Cups (Coarse) 12pkFDAComposite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment.
  6. Olympus Corporation of the Americas recalls Brand NameFDAFollowing a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may…
  7. PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls ALLURA Xper FD20 Biplane and 21 moreFDAPhilips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
  8. Baxter Healthcare Corporation recalls Blue Ventilator Adapter ModuleFDAUrgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of…
  9. Synthes (USA) Products LLC recalls .4 VOLT TAP LOCKING SCREW TAP TO 90MM and 1 moreFDAA full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps…
  10. BD KIESTRA LAB AUTOMATION recalls BD Kiestra" ReadAFDAIn certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this…
  11. Mentor Texas, LP. recalls Artoura Breast Tissue Expanders Reference Numbers and 1 moreFDAInfusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
  12. SUMMA THERAPEUTICS, LLC recalls Brand NameFDAPotential for the balloon in the device to not meet burst specifications.
  13. Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E)FDAA potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP…
  14. Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 18G (for Aurora Trackers) and 4 moreFDAThere was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip…
  15. Stryker Communications recalls Stryker CHROMOPHARE Softlit Ring Surgical Light System REFFDASurgical light assembly may not adequate support the weight of the ceiling cover.
  16. Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000FDADue to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll…
  17. PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Vue Motion V12FDAPotential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
  18. Medline Industries, LP recalls Medline Kits containing ENFIT G-TUBE CONNECTOR 1 and 1 moreFDAConnectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to…
  19. Medline Industries, LP recalls Sterile Radiology Procedure KitsFDAThere is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.
  20. Medline Industries, LP recalls THORACIC ROBOTS and 4 moreFDAKits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the…
  21. Cardinal Health 200, LLC recalls Cardinal Health Monoject Tuberculin SyringeFDAThe outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the…
  22. Oxoid Australia Pty Limited recalls OXOID MICROBACT IDENTIFICATION KITSFDAOxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to…
  23. Philips Respironics, Inc. recalls Philips Respironics Trilogy Evo O2 and 3 moreFDAIn some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
  24. GE Healthcare LLC recalls SIGNA Premier systemsFDAWhole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with…
  25. Dimension® Creatinine (CRE2)Health CanadaPerformance
  26. Edwards EVOQUE Tricuspid Delivery SystemHealth CanadaLabelling and packaging
  27. AneurysmFlowHealth CanadaPerformance
  28. Olympus Connecting TubesHealth CanadaPerformance
  29. Ultraview SL™ Command ModuleHealth CanadaPerformance
  30. Flamingo Funnel™Health CanadaSterility
  31. DermaHealer UVB Light Therapy LampsHealth CanadaUnauthorised device
  32. Kiwi® Vacuum Assisted Delivery SystemHealth CanadaPerformance
  33. Philips Ultrasound, Inc recalls Philips S5-2 Ultrasound Transducer and 10 moreFDATo provide clarification and labelling to define the useful life of ultrasound transducers in the field.
  34. CareFusion 303, Inc. recalls BD Alaris" System with Guardrails" Suite MX with Point of Care UnitFDADue to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi…
  35. I.T.S. GmbH recalls HCS and 11 moreFDAUpdated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
  36. Olympus Corporation of the Americas recalls Olympus Endoscope Suction Valve and 3 moreFDAThe MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
  37. AMO Puerto Rico Manufacturing, Inc. recalls Brand NameFDADue to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding…
  38. Erbe USA Inc recalls Flexible Cryoprobe (OD 1.1mm and 2 moreFDAProbes may rupture/burst during activation
  39. Waldemar Link GmbH & Co. KG (Mfg Site) recalls Embrace Drill TowerFDAThe surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However,…
  40. Medtronic MiniMed, Inc. recalls MiniMed" 780GFDAThree software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions…
  41. GE Medical Systems, LLC recalls Imactis CT-Navigation SystemFDAThere is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip…
  42. Navajo Manufacturing Company recalls Handy Solutions Neck & Shoulders Heating PadFDAIf heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by…
  43. Olympus Corporation of the Americas recalls Brand NameFDAOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective…
  44. Merit Medical Systems, Inc. recalls CentrosFLO Hemodialysis Catheters and 4 moreFDA16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies,…
  45. Tandem Diabetes Care, Inc. recalls Tandem pumps are battery-operated infusion pumps capable of both ba…FDAThe Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify…
  46. Medline Industries, LP recalls Puracol and Puracol Plus Collagen Wound DressingsFDAElevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation,…
  47. Orthalign, Inc recalls Drill PlateFDADrill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
  48. Philips Respironics, Inc. recalls Philips Respironics Trilogy Evo Universal and 3 moreFDAUsing non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume…
  49. Beckman Coulter, Inc. recalls MicroScan Neg MIC 56 REF C42464 UDI-DI codeFDADue to the likely presence of contamination in well(s).
  50. Koven Technology, Inc. recalls Brand NameFDAProduct labeling includes a fetal indication for use that is not cleared under its 510(k).