Medical devices recalls
2,678 recalls since January 2025 from Health Canada, FDA, newest first. Last notice 19 Nov 2026.
Page 23 of 54
- Philips IntelliVue Patient MonitorsHealth CanadaPerformance
- Alaris & BD Alaris Point-Of-Care Unit (PCU)Health CanadaDevice compatibility
- Celon ProBreath and Celon ProSleep ApplicatorsHealth CanadaLabelling and packaging
- BD Alaris™ Syringe PumpsHealth CanadaPerformance
- BioFlo Dialysis Catheter With Endexo TechnologyHealth CanadaPerformance
- Bipolar Hugo RAS SystemsHealth CanadaPerformance
- Monarch PlatformHealth CanadaPerformance
- Trilogy EVO and EV300Health CanadaDevice compatibility
- Raha Generator ShieldHealth CanadaPerformance
- Kiestra ReadAHealth CanadaPerformance
- Universal Viewer - System FeatureHealth CanadaPerformance
- Leksell Gamma KnivesHealth CanadaPerformance
- Laerdal Medical Corporation recalls Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4FDAUnits manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA…
- Hologic, Inc recalls Brevera Breast Biopsy System Disposable 9 Gauge NeedleFDAPresence of particulates in affected devices that can be deposited into breast tissue during use.
- Intuitive Surgical, Inc. recalls da Vinci 5 Surgeon Console Viewer DisplayFDARobotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field…
- Elekta, Inc. recalls Brand NameFDADue to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan…
- Philips North America Llc recalls Brilliance iCT and 3 moreFDAAs part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of…
- Alphatec Spine, Inc. recalls ATEC Lateral Navigation Disc Prep Instruments LIF NAVFDADue a design issue where the navigated array connection geometry is incorrect.
- Straumann USA LLC recalls WB Impression Post Closed Tray inclFDAThe mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown.
- Steris Corporation recalls Barco MNA with the HexaVue IP Integration SystemFDAThis is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter)…
- Medline Industries, LP recalls Reprocessed Abbott Inquiry Steerable Diagnostic Catheters and 6 moreFDAThese lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory…
- Medline Industries, LP recalls Medline medical convenience kits packaged as and 1 moreFDAMedline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD…
- Medline Industries, LP recalls Medline 14 Fr Suction Catheter Kits SKU DYND40972 and 1 moreFDAA smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use…
- SUREPULSE MEDICAL LTD recalls Surepulse VS Cap Large REF and 4 moreFDARetroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
- AVID Medical, Inc. recalls Halyard TRANSPORT BAG KITFDATyvek bag seal issue which may compromise sterility of the kit.
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M12 System Model Numbers and 7 moreFDAUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Erbe Flexible CryoprobeHealth CanadaPerformance
- Medline Kits containing Surgical Gowns, Packs and DrapesHealth CanadaSterility
- Medihoney® Wound and Burn ProductsHealth CanadaSterility
- Changchun Wancheng Bio-Electron Co., Ltd. recalls Male Fertility Sperm Test for Home Use (Cassette) and 10 moreFDATest strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S.…
- Beckman Coulter, Inc. recalls COULTER DxH DiluentFDADue to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
- Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump Catalog NumbersFDAA software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting…
- Olympus Corporation of the Americas recalls Brand NameFDAA deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
- Intuitive Surgical, Inc. recalls Brand NameFDADue to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was…
- ETAC A/S recalls Brand NameFDAA customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation…
- Tornier, Inc recalls TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILEFDAA specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during…
- FUJIFILM Healthcare Americas Corporation recalls FDR Visionary SuiteFDAIt was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the…
- Focalyx Technologies, LLC. recalls Focalyx FusionFDASoftware device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not…
- Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubingFDAInternal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor.…
- Encore Medical, LP recalls EMPOWR 3D Knee Tibial Insert 5R 14MM and 1 moreFDA942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices…
- VANTIVE US HEALTHCARE LLC recalls PRIMSAFLEX M100 SET and 6 moreFDAThere is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- Siemens Medical Solutions USA, Inc recalls NAEOTOM Alpha.Pro Software applications and 3 moreFDATo remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
- Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation SystemFDASample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or…
- Fujirebio Diagnostics, Inc. recalls Product NameFDACustomers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
- Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm and 3 moreFDAIncreased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the…
- Olympus Corporation of the Americas recalls Product Name and 1 moreFDADevices which did not undergo thermoforming could deform and lose performance.
- Vortex Surgical Inc. recalls Convenience Kits 1 and 12 moreFDAThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has…
- Biomerieux Inc recalls VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B…FDAPotential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae…
- Wilson-Cook Medical Inc. recalls Product NameFDANasal feeding tube packaged without the nasal transfer tube component.
- VISERA™ Imaging SystemHealth CanadaPerformance