Dexcom, Inc. recalls Dexcom G6
- Published
- 26 Feb 2025
- Units affected
- 3 products
- Agency
- FDA
- Country
- United States
The hazard
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OR-001
Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001
- Brand / company
- Dexcom, Inc.
- Codes / lots
- All Serial Numbers/UDI: 00386270000590 | All Serial Numbers/UDI: 00386270000583 | All Serial Numbers/UDI: 00386270001900
- Distribution
- Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC, NH, FL, CA, NJ, PR, GA and the countries of Austria, Belgium, Canada, Switzerland, Germany, Spain, France, Ireland, Japan, Netherlands, Slovenia, United Kingdom, Australia, Czech Republic, Denmark, Finland, France, Hong Kong, Israel, Italy, Korea, Kuwait, New Zealand, Norway, Oman, Poland, Qatar, Saudi Arabia, Singapore, Slovakia, South Africa, Sweden, Turkey, United Arab Emirates.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 26 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.