Dexcom, Inc. recalls Dexcom G7 iOS Application
- Published
- 11 Jun 2025
- Units affected
- 61,355 users
- Agency
- FDA
- Country
- United States
The hazard
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
- Brand / company
- Dexcom, Inc.
- Codes / lots
- Software version 2.7.0/ UDI: 00386270001863
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR), Kuwait (KW), Lebanon (LB), Lithuania (LT), Luxemburg (LU), Malta (MT), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Poland (PL), Portugal (PT), Qatar (QA), Saudi Arabia (SA), Sweden (SE), Slovenia (SI), Slovakia (SK), Turkey (TR) and South Africa (ZA).
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.