The hazard
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
ICU Medical Plum Solos, Item number 400011001; infusion pump
ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump
- Brand / company
- ICU Medical, Inc.
- Codes / lots
- UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.
- Recall status
- Ongoing