Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe Pump

Read the official FDA notice

Published
18 Jun 2025
Units affected
8027 units
Agency
FDA
Country
United States

The hazard

Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Baxter Novum IQ Syringe Pump, product code 40800BAXUS,

Brand / company
Baxter Healthcare Corporation
Codes / lots
UDI/DI 05413765852428, All Serial Numbers
Distribution
US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 18 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.