Beckman Coulter, Inc. recalls Access Testosterone assay
- Published
- 11 Jun 2025
- Units affected
- 70 units
- Agency
- FDA
- Country
- United States
The hazard
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Access Testosterone assay, Catalog Number 33560
- Brand / company
- Beckman Coulter, Inc.
- Codes / lots
- UDI-DI: 15099590231347; Serial Number: 538081
- Distribution
- US State: FL. UAE, Vietnam
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.