CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

ICU Medical, Inc. recalls ICU Medical ChemoLock Closed Vial Spike…

ICU Medical, Inc. recalls ICU Medical ChemoLock Closed Vial Spike w/Skirt and 35 more

Read the official FDA notice

Published
5 Nov 2025
Units affected
38 products
Agency
FDA
Country
United States

The hazard

Port weld of drug transfer device may separate or break during use and potentially result in a leak.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5

ICU Medical ChemoLock Port w/Spiros, Red Cap REF: CL-35

ICU Medical ChemoLock" Port, Bulk, Non-Sterile REF: CL2100-NS

ICU Medical ChemosafeLock Connecter REF: KL-FNU3

ICU Medical Syringe Transfer Set w/MicroClave, ChemoLock Port REF: CL-34

ICU Medical ChemoLock Vented Vial Spike, 20mm, 20 Units ¿ REF: 011-CL-70S-20

ICU Medical Chemo Lock Universal Vented Vial Spike REF: CL-70-10

ICU Medical 21" Ext Set w/ChemoLock Port, Clamp, Graduated Connector REF: CL3360

ICU Medical 6.5" Bag Spike w/ChemoLock Port Additive Port, ChemoLock Port REF: CL3955

ICU Medical 30" (76 cm) Appx 3.3 mL, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros", Hanger, Drop-In Red Cap REF: CL3011

ICU Medical ChemosafeLock Vial Adapter REFs: KL-VA001U3 KL-VA002U3 KL-VA131U3 KL-VA201U3 KL-VA202U3 KL-VA321U3

ICU Medical Oncology Transfer Kit w/ChemoClave and ChemoLock Port; Spiros w/Red Cap; ChemoLock Port REF: CL4138

Brand / company
ICU Medical, Inc.
Codes / lots
UDI:10887709077116/Lot:14395313, 14393195, 14396054 | Lot:14407070,14384996,14391949; Distributed in Canada | UDI:00887709094505/Lot: 14407400 | Lot:14399518, 14399521, 14406408, 14399520, 14406410, 14406411 ; Distributed in Japan | Lot:14400199; Distributed in Canada | Lot:14401659/ Distributed in France
Distribution
Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY and OUS (foreign) to countries of: Canada, France, and Japan.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 5 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.