The hazard
It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
- Brand / company
- Canon Medical System, USA, INC.
- Codes / lots
- Serial Numbers: SZA2422001, SXA2432001, SUC23Z2014, BBF2392003, SUC2392013, SZA2422002, SXA2432002, SZA2442004.
- Distribution
- U.S.
- Recall status
- Ongoing