The hazard
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
- Brand / company
- Canon Medical System, USA, INC.
- Codes / lots
- CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.
- Distribution
- US distribution to Ohio and Nevada
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.