The hazard
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
- Brand / company
- ICU Medical, Inc.
- Codes / lots
- LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227
- Distribution
- US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
- Recall status
- Ongoing