CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

GE Medical Systems, LLC recalls Allia IGS 3 Pulse angiographic X-ray…

GE Medical Systems, LLC recalls Allia IGS 3 Pulse angiographic X-ray system and 4 more

Read the official FDA notice

Published
17 Jun 2026
Units affected
5 products
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Allia IGS 3 Pulse angiographic X-ray system

Allia IGS Pulse angiographic X-ray system

Allia Moveo angiographic X-ray system

Allia IGS 5 Pulse angiographic X-ray system

Allia IGS 7 Pulse angiographic X-ray system

Brand / company
GE Medical Systems, LLC
Codes / lots
UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030 | UDI-DI: 00195278827647; Serial Numbers: DVMSS2500001HL, DVMSS2500002HL, DVMSS2500003HL, DVMSS2500004HL, DVMSS2500005HL, DVMSS2500006HL, DVMSS2600001HL, DVMSS2600002HL | UDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005 | UDI-DI: 00195278719263; Serial Numbers: M2-23-039, M2-23-056, M2-24-004, M2-24-019, M2-24-020, M2-24-025, M2-24-030, M2-24-031, M2-24-032, M2-24-033, M2-24-034, M2-24-037, M2-24-038, M2-24-039, M2-24-040, M2-24-046, M2-24-057, M2-24-062, M2-24-064, M2-24-066, M2-24-069, M2-24-071, M2-24-077, M2-24-078, M2-24-080, M2-24-083, M2-25-003, M2-25-004, M2-25-005, M2-25-011, M2-25-013, M2-25-015, M2-25-017, M2-25-019, M2-25-023, M2-25-024, M2-25-025, M2-25-026, M2-25-028, M2-25-032, M2-25-036, M2-25-037, M2-25-039, M2-25-040, M2-25-041, M2-25-042, M2-25-043, M2-25-046, M2-25-048, M2-25-049, M2-25-056, M2-25-057, M2-25-058, M2-25-062, M2-25-065, M2-25-067, M2-25-072, M2-25-074, M2-25-077, M2-25-078, M2-25-080, M2-25-250, M2-25-300, M2-26-002, M2-26-004, M2-26-006, M2-26-0
Distribution
Worldwide - US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 17 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.