The hazard
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Optima XR200amx, Mobile Digital-Ready Radiographic System
Optima XR220amx, Mobile Digital-Ready Radiographic System
Optima XR240amx, Mobile Digital Radiographic System
- Brand / company
- GE Medical Systems, LLC
- Codes / lots
- UDI-DI: 00840682115278; System IDs: 419383DP3, 773792XR200, 856355OPT, FEB520100 | UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4 | UDI-DI: 00840682122351 and 00840682123686; System IDs: 252847PP249, 281MSL2402, 291OPTI1, 312CCHSUPER7, 318487OPTIMA1, 360744OPT2402, 501686OPTI1, 504826XR241, 519941XR240A, 540776ODP3, 541472M4, 602344240C, 706291BEE9, 713799XR240, 770836BEE2, 813872OPT11, 813872OPT9, 843724OPT1, 843847OPT240, 870239OPT2, 905522AMX240, 973625DIG4, DF2402001002WK, SA1013MM13
- Distribution
- Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.