The hazard
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Brand / company
- Baxter Healthcare Corporation
- Codes / lots
- UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 8 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.