CheckMyRecalls

10,013 recalls · updated 20 Nov 2026

GE Medical Systems, LLC recalls GE HealthCare OEC 9900 Elite C-arm…

GE Medical Systems, LLC recalls GE HealthCare OEC 9900 Elite C-arm Systems. and 1 more

Read the official FDA notice

Published
3 Sept 2025
Units affected
2 products
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GE HealthCare OEC 9900 Elite C-arm Systems.

GE HealthCare OEC 9800.

Brand / company
GE Medical Systems, LLC
Codes / lots
GTIN: 00840682114349 | GTIN: 00840682114349
Distribution
US Nationwide distribution.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 3 Sept 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.